The Viability Dyes can be used to exclude non-viable cells in flow cytometry assays. They irreversibly bind to cell su rface and intracellular amines. The dye can covalently bind to a higher concentration of amines in cells that have lost plasma membrane integrity, such as those in the late stages of apoptosis or dead cells, and generate a significantly more intense fluore scence signal. Dead d ells stained with the viability dye will maintain their staining intensity after washing, fixation, permeabilization, or additional intracellular staining.
Protocol
1.Allow vial of Viability Dye to equilibrate to room temperature and quickly spin before use.
2.Wash cells twice in PBS solution free of a zide, serum or protein.
3.Resuspend cells in azide,serum, and protein free PBS at a concentration of 1-10x106/mL.
4.Add 1 μL of Fixable Viability Dye per 1 mL of cells and vortex immediately.
5.Incubate the cells for 30 minutes protected from light at 2-8 °C.
6.Wash the cells with flow staining buffer or equivalent.
7.Cells are ready to be stain, fixed, and permeabilized as desired.
Da — when not otherwise indicated, molecular weight units are daltons. Mw — weight-average molecular weight. Mn — number-average molecular weight.
Products are supplied for research and development use only. Not for use in humans, animals, diagnosis, or therapy.
Cite this article
Aladdin Scientific. "Staining Dead Cells with Viability Dyes" Aladdin Knowledge Base, updated 9 juin 2025. https://www.aladdinsci.com/us_fr/faqs/staining-dead-cells-with-viability-dyes-en.html
Was this article helpful?YesNo8 out 17 found this helpful
Nous utilisons des cookies pour garantir le bon fonctionnement du site et, lorsque cela est autorisé, pour améliorer votre expérience. Vous pouvez gérer vos préférences à tout moment dans les Paramètres. Pour en savoir plus, consultez notre Politique relative aux cookies.
Souhaitez-vous que nous vous envoyions un message lorsque des réductions seront disponibles?
Me rappeler plus tard
Merci ! Veuillez vérifier votre boîte de réception e-mail pour confirmer.
Products are supplied to verified businesses, institutions, and qualified professionals for research and development use only. Not for use in humans, animals, diagnosis, or therapy.