CALIDAD & PUREZA BioReagent ? BioReagent grade — tested suitable for life-science and molecular-biology use. Use for cell culture, assays, and biochemical work needing biological compatibility. for Enzyme immunoassay(ELISA) ? ELISA grade — low-background reagents validated for enzyme immunoassays. Use to build sensitive, reproducible ELISA assays.
★
Talla
Alemania (EU)
USA*
Precio
Cantidad
48T
EJ1515243-48T
Fabricado bajo pedido · 8–12 semanas
563,94€
96T
EJ1515243-96T
Fabricado bajo pedido · 8–12 semanas
841,62€
Enter a quantity for the sizes you want to add.

Descripción general

Vomitoxin, also known as Deoxynivalenol (DON), is a B-type trichothecene and an epoxy compound. This mycotoxin mainly occurs in cereals such as wheat, barley, oats, rye and corn, while it is less frequently found in rice, sorghum and clover. The formation of deoxynivalenol is primarily associated with Gibberella zeae and F. culmorum, both important plant pathogens. They cause Fusarium head blight in wheat and Fusarium ear rot in corn. The incidence of Fusarium head blight is closely related to the humidity during flowering (anthesis); the timing of rainfall, rather than rainfall amount, is the most critical factor. However, due to toxin leaching, increased moisture close to harvest is associated with lower vomitoxin content in wheat kernels.

Principle:

This kit adopts competitive enzyme-linked immunosorbent assay (ELISA). Add samples, standards and Biotin-labeled antibody, followed by HRP conjugate sequentially into microplates pre-coated with vomitoxin antigen. Incubate and wash between steps. Develop color with TMB substrate. TMB turns blue under the catalysis of horseradish peroxidase (HRP), and finally turns yellow after acid termination. The color intensity is negatively correlated with the concentration of vomitoxin in the sample. Measure the absorbance (OD value) at 450 nm with a microplate reader, and calculate the concentration of the sample.

Precautions

1. Strictly follow the specified incubation time and temperature to guarantee accurate results. All reagents must be equilibrated to room temperature (20–25°C) before use. Refrigerate reagents immediately after use.

2. Improper plate washing may lead to inaccurate readings. Ensure complete removal of residual liquid in wells prior to substrate addition. Avoid prolonged drying of microwells throughout the procedure.

3. Wipe residual liquid and fingerprints from the bottom of the microplate, as contaminants will interfere with OD values.

4. The TMB substrate solution shall remain colorless; discard substrate solution that has turned blue.

5. Prevent cross-contamination between reagents and samples to avoid erroneous test results.

6. Avoid direct intense light exposure during reagent storage and incubation.

7. All reaction reagents must not come into contact with bleach solvents or their volatile fumes; bleach components will destroy the biological activity of kit reagents.

8. Do not use expired products; components from different lot numbers or batch codes cannot be mixed.

9. Recombinant proteins from external sources may fail to be recognized due to mismatched antibody affinity with kit capture antibodies.

10. Handle all potentially infectious samples properly, and dispose of specimens and testing devices in accordance with standardized biosafety protocols.

Kit Components:

Component No.Component NameAppearance48T96TStorage
EJ1515243APre-coated Assay Plate—48 wells96 wells2-8℃
EJ1515243BStandardSolid1 vial2 vials2-8℃
EJ1515243CUniversal DiluentLiquid1×20mL2×20mL2-8℃
EJ1515243DBiotin-antibody (100×)Liquid30μL60μL2-8℃
EJ1515243EStreptavidin-HRP (100×)Liquid60μL120μL2-8℃
EJ1515243FWash Buffer (20×)Liquid1×10mL2×10mL2-8℃
EJ1515243GTMB SubstrateLiquid5mL10mL2-8℃
EJ1515243HStop SolutionLiquid3mL6mL2-8℃
EJ1515243IPlate Sealer—4 pieces4 pieces2-8℃
Note: EJ1515243B, EJ1515243D, EJ1515243E and EJ1515243F shall be diluted in accordance with the instructions.

Sample Handling & Requirements

1. The detection range of the kit is not equivalent to the concentration range of the analyte in the sample. It is recommended to estimate the analyte concentration in the sample by relevant literature before the experiment and confirm the actual concentration of the sample through pre-experiment. If the concentration of the analyte in the sample is too high or too low, appropriately dilute or concentrate the sample.

2. If the test sample is not among the sample types listed in the manual, a pre-experiment is recommended to verify the detection validity.

3. Solid samples (cereals, feed, nuts, etc.): ① Weigh the sample. Crush large particles, and sieve the crushed sample through a 1–2 mm test sieve. Weigh 2.0 g into a centrifuge tube, add 20 mL of 1% NaCl (prepared in 70% methanol solution), shake for 5 min, and centrifuge at 5000 r/min for 10 min. ② Dilute the supernatant 1:1 with 1% NaCl (prepared with distilled water). For example, mix 100 μL supernatant with 100 μL 1% NaCl solution, and mix well for testing.

4. Sample appearance: The sample shall be clear and transparent; suspended solids shall be removed by centrifugation.

5. Sample storage: Samples can be stored at 4 ℃ if tested within 1 week after collection. If testing cannot be performed promptly, aliquot the sample into single-use portions and store frozen at -20 ℃ (test within 1 month) or -80 ℃ (test within 6 months). Avoid repeated freeze-thaw cycles. Hemolysis will interfere with the final test result, so hemolyzed samples are not suitable for this assay.

Sample Dilution Protocols 

Please estimate the concentration range of the sample in advance. If your test sample requires dilution, please refer to the dilution protocol below: 

100-fold dilution: One-step dilution. Pipette 5 μL sample into 495 μL Universal Diluent for 100-fold dilution; 

1000-fold dilution: Two-step dilution. Pipette 5 μL sample into 95 μL Universal Diluent for 20-fold dilution. Then pipette 5 μL of the 20-fold diluted sample into 245 μL Universal Diluent for 50-fold dilution, for a total dilution of 1000-fold; 

100000-fold dilution: Three-step dilution. Pipette 5 μL sample into 195 μL Universal Diluent for 40-fold dilution. Then pipette 5 μL of the 40-fold diluted sample into 245 μL Universal Diluent for 50-fold dilution. Finally, pipette 5 μL of the 2000-fold diluted sample into 245 μL Universal Diluent for 50-fold dilution, for a total dilution of 100000-fold; 

The volume pipetted at each dilution step shall be no less than 3 μL, and the dilution factor per step shall not exceed 100-fold. Mix thoroughly after each dilution step and avoid bubble formation. 

Required Lab Equipment (User-Supplied)

1. Microplate reader (capable of 450 nm absorbance measurement)

2. High-precision pipettes and matching tips: 0.5–10 μL, 5–50 μL, 20–200 μL, 200–1000 μL

3. 37°C incubator

4. Distilled or deionized water

Pre-Assay Reagent Preparation

1. Take the kit out of the refrigerator 10 minutes in advance and equilibrate it to room temperature.

2. Preparation of standard serial working solutions: Add 1 mL Universal Diluent to the lyophilized standard. Let it stand for 15 minutes for complete dissolution, then mix gently (the concentration is 200 ng/mL). Prepare serial dilutions to the following concentrations: 200 ng/mL, 100 ng/mL, 50 ng/mL, 25 ng/mL, 12.5 ng/mL, 6.25 ng/mL, 3.125 ng/mL, 0 ng/mL.

Serial dilution method: Prepare 7 EP tubes, and add 500 μL Universal Diluent to each tube. Pipette 500 μL of the 200 ng/mL standard working solution into the first EP tube and mix well to obtain the 100 ng/mL standard working solution. Repeat this pipetting and mixing step sequentially for the remaining tubes. The last tube serves directly as the blank control; no liquid is transferred from the second-to-last tube. See the figure below for details.


3. Preparation of Biotin-antibody Working Solution: Centrifuge the 100× concentrated Biotin-antibody at 1000×g for 1 min, 15 min prior to use. Dilute the 100× concentrated Biotin-antibody to 1× working concentration with Universal Diluent (e.g., 10 μL concentrate + 990 μL Universal Diluent). Prepare freshly before use.

4. Preparation of Conjugate Working Solution: Centrifuge the 100× concentrated conjugate at 1000×g for 1 min, 15 min prior to use. Dilute the 100× concentrated conjugate to 1× working concentration with Universal Diluent (e.g., 10 μL concentrate + 990 μL Universal Diluent). Prepare freshly before use.

5. Preparation of 1× Wash Buffer: Add 10 mL of 20× Wash Buffer into 190 mL distilled water. (Crystallization may occur in concentrated wash buffer taken out from the refrigerator, which is normal. Place it at room temperature until crystals are fully dissolved before preparation.)

Assay Procedure

1. Take out the required strips from the aluminum foil bag which has been equilibrated at room temperature for 10 minutes. Seal the remaining strips in a self-sealing bag and store at 4℃.

2. Sample loading: Add 50 μL of samples or standards of different concentrations to the corresponding wells. Add 50 μL Universal Diluent to the blank wells, then add 50 μL Biotin-antibody working solution to each well. Cover with a Plate Sealer and incubate at 37℃ for 60 min. (Recommendation: Dilute the test sample at least 1-fold with Universal Diluent before loading to reduce matrix effect on test results. Multiply the measured concentration by the corresponding dilution factor during final calculation. It is recommended to set replicate wells for all test samples and standards.)

3. Washing: Discard the liquid. Add 300 μL of 1× Wash Buffer to each well, let it stand for 1 min, flick out the wash buffer and pat dry on absorbent paper. Repeat the washing process 3 times (a plate washer can also be used).

4. Addition of conjugate working solution: Add 100 μL conjugate working solution to each well. Cover with a Plate Sealer and incubate at 37℃ for 30 min.

5. Washing: Discard the liquid and wash the plate 5 times following the washing procedure in Step 3.

6. Addition of substrate: Add 90 μL TMB Substrate to each well. Cover with a Plate Sealer and incubate at 37℃ in the dark for 15 min.

7. Addition of Stop Solution: Take out the plate, add 50 μL Stop Solution directly to each well, and immediately measure the OD value of each well at 450 nm.

Calculation of Test Results: 

Result Interpretation: 

1. Calculate the average OD values of replicate wells for standards and samples. Plot the four-parameter logistic standard curve on a log-log graph with concentration as the X-axis and OD value as the Y-axis (exclude the blank group when plotting).

2. If the OD value of the sample is higher than the upper limit of the standard curve, dilute the sample appropriately and re-test. Multiply the calculated concentration by the corresponding dilution factor.

Typical Data and Reference Curve:

The following data and curve are for reference only. Users shall establish the standard curve according to their own experiments.

Concentration (ng/mL)200100502512.56.253.1250
OD Value0.2160.2980.4170.6690.9551.4051.791.865

Note: This figure is for reference only. Sample concentrations shall be calculated using the standard curve generated from data of each individual experiment.

Kit Performance

1. Precision: Intra‑assay coefficient of variation <10%; inter‑assay coefficient of variation <10%.

2. Recovery: Vomitoxin at three different concentration levels was spiked into selected healthy serum, plasma and cell culture supernatant for recovery calculation.

Sample TypeRange (%)Mean Recovery (%)
Serum (n=8)81-10196
Plasma (n=8)92-105101
Cell Culture Supernatant (n=8)97-108105

3. Linearity‑of‑dilution: High-concentration Vomitoxin was spiked into 4 selected healthy serum, plasma and cell culture supernatant samples respectively. Dilutions were performed within the dynamic range of the standard curve to assess linearity.

Dilution RatioRecovery (%)SerumPlasmaCell Culture Supernatant
1:2Range83-9588-9690-110
1:2
Mean Recovery919396
1:4Range89-10487-108105-115
1:4Mean Recovery9398110

Troubleshooting Guide

If unsatisfactory assay results occur, take photos of color development, save experimental data, retain used microplate strips and unused reagents, then contact our technical support for troubleshooting. Refer to the troubleshooting items below:

1. Poor standard curve linearity 

① Incorrect standard dilution: Reconstitute and dilute standard strictly following recommended protocol. 

② Inaccurate pipetting: Calibrate pipettes regularly and verify tip tightness/sealing performance. 

③ Evaporation of reaction solution: Seal microplate with plate sealer film. 

④ Incomplete washing: Perform sufficient wash cycles with adequate volume of wash buffer. 

⑤ Foreign residues at the bottom of wells: Wipe the bottom of the microplate before absorbance reading.

2. Faint or absent color development 

① Insufficient incubation duration: Adhere to specified incubation time. 

② Incorrect incubation temperature: Incubate at recommended temperature. 

③ Insufficient reagent addition volume: Verify pipette accuracy and follow operating procedure strictly. 

④ Improper reagent dilution: Double-check dilution steps for all reagents. 

⑤ Inactivated Streptavidin-HRP: Mix Streptavidin-HRP with substrate and verify activity via color reaction test.

3. Low OD readings 

① Wrong microplate reader parameter: Confirm detection wavelength setting on instrument. 

② Missing stop solution addition: Add appropriate volume of stop solution. 

③ Overlong delay before plate reading: Measure absorbance immediately after termination. 

④ Excessively high analyte concentration in sample: Determine optimal dilution factor via preliminary pre-test. 

⑤ Too low analyte concentration in sample: Determine optimal dilution factor via preliminary pre-test.

4. High background signal 

① Contaminated TMB substrate: Replace with fresh substrate solution. 

② Excess substrate incubation time: Strictly control color development duration. 

③ Wrong dilution for detection antibody or Streptavidin-HRP: Dilute reagents in accordance with official recommended protocol. 

④ Incomplete plate washing: Carry out adequate wash cycles with sufficient wash buffer volume.

Especificaciones

Especificaciones y pureza
BioReagent, for Enzyme immunoassay(ELISA)
Grado
BioReagent, para el inmunoensayo enzimático (ELISA)
Aplicación
ELISA
Tipo de muestra
Rango de detección
3.125ng/mL~200ng/mL
Especificidad
Detecting Vomitoxin in samples, with no significant cross-reactivity to other homologs.
Detection instrument
Microplate reader
Detection wavelength
450nm
Repeatability
The intra- and inter-assay coefficients of variation (CV) are both <10%.
Detection duration
3.5 hours, not including equilibration and sample preparation time.
Almacenamiento y envío
Condiciones de almacenamiento de almacenamiento
Store at 2-8°C
Enviado en
Wet ice
Estabilidad y almacenamiento
Store at 2-8℃ long term (6 months).
Contents & Storage
Component No.Component NameAppearance48T96TStorage
EJ1515243APre-coated Assay Plate—48 wells96 wells2-8℃
EJ1515243BStandardSolid1 vial2 vials2-8℃
EJ1515243CUniversal DiluentLiquid1×20mL2×20mL2-8℃
EJ1515243DBiotin-antibody (100×)Liquid30μL60μL2-8℃
EJ1515243EStreptavidin-HRP (100×)Liquid60μL120μL2-8℃
EJ1515243FWash Buffer (20×)Liquid1×10mL2×10mL2-8℃
EJ1515243GTMB SubstrateLiquid5mL10mL2-8℃
EJ1515243HStop SolutionLiquid3mL6mL2-8℃
EJ1515243IPlate Sealer—4 pieces4 pieces2-8℃
Note: EJ1515243B, EJ1515243D, EJ1515243E and EJ1515243F shall be diluted in accordance with the instructions.

Datos avanzados

Certificados (CoA, COO, BSE/TSE y tabla de análisis)
C of A & Other Certificates(BSE/TSE, COO):
Analytical Chart:
Preguntas frecuentes y artículos
Calculadoras de soluciones
Reseñas

Reseñas de cliente

Preguntas frecuentes

What does BioReagent mean?
BioReagent indicates that the product has been processed and tested for biological and biochemical applications. Specifications may include endotoxin limits, microbial controls, sterility, low DNase/RNase/protease activity, and verified biological activity where relevant.
How should this product be stored?
Store at 2–8 °C. Refrigerated storage is required to maintain the specified shelf life.
How is this product shipped?
This product ships chilled on wet ice. Unpack on arrival and transfer it to the storage condition stated above.
What documentation is provided?
Available product documentation, including Certificates of Analysis (COA), Safety Data Sheets (SDS), and specification sheets, is shown in the product document area. Document availability and access follow the current site policy.

¿Necesita ayuda para elegir la calidad?

Nuestra guía de selección de calidades cubre la pureza, el estado del estabilizador y la idoneidad de la aplicación para todas las variantes de nuestro catálogo.

Ver la guía de calidades BioReagent → Ver la guía de calidades for Enzyme immunoassay(ELISA) →

¿Quieres que te enviemos un mensaje cuando tengamos descuentos disponibles?

Recordármelo más tarde

¡Gracias! Por favor, revisa tu bandeja de entrada de correo electrónico para confirmar.

¡Vaya! Las notificaciones están desactivadas.