This lab was derived from: Animal Cell Culture - A Guide to Basic Techniques (5th Edition)

Operation method

Programme 11.4 Sterilization of filtration units

Materials and Instruments

Filter Housings Shelves Prefilters Fiberglass Nylon Bags Autoclave Sterility Indicator Tape Sterilization Kettles

Move

1. After a thorough cleaning with detergent, rinse the filter unit with water, then with deionized water and allow to air dry.

2. Insert a filter plate into the filter and place the filter membrane onto the plate, wetting it to eliminate static electricity if the membrane is made of polycarbonate.

3. place the pre-filter (fiberglass and other desired substances) on top of the filter.

4 . Install the filter bracket, do not tighten completely (you can leave a screw ring).

5. seal the inlet and outlet of the filter with aluminum foil.

6. Wrap the filter unit in sterilized paper or steam-permeable nylon film and apply sterile indicator tape.

7. High-pressure sterilize the filter at 121°C, 100 kPa (1 bar, 15 lb/in2 ) for 20 min without the pre-vacuum and post-vacuum cycles (when using an automatic autoclave, use the "Liquid Cycle" procedure).

8. Remove the filter and cool it.

9. Do not tighten the filter holder completely until filtration has begun and the membrane is already wet.

Caveat

1. Selection of sterilization methodsMany plastic utensils cannot be autoclaved or dry heat sterilized because the temperature is too high. When sterilizing these vessels, soak them in 70% ethanol for 30 minutes, then place them in a laminar flow hood and dry them under ultraviolet light. Particular care should be taken with some plasticware (e.g., Plexiglas, plastic glass, clear synthetic resins) because they will depolymerize in ethanol or when exposed to UV light. Vinyl oxide can be used to sterilize plasticware, but it takes two to three weeks after sterilization to wash it all off the plastic surface. The best method of sterilizing plasticware is y-irradiation, standardized to 25 kGy. Ware should be packaged and sealed in polyethylene plastic with a heat-welded closure.


For more product details, please visit Aladdin Scientific website.

https://www.aladdinsci.com/

Categories: Protocols
Explore topics: Cellular experiment

Da — when not otherwise indicated, molecular weight units are daltons.   Mw — weight-average molecular weight.   Mn — number-average molecular weight.

Products are supplied for research and development use only. Not for use in humans, animals, diagnosis, or therapy.

Cite this article

Aladdin Scientific. "Sterilization of filtration units" Aladdin Knowledge Base, updated 24 dic 2024. https://www.aladdinsci.com/us_es/faqs/sterilization-of-filtration-units-en.html
Was this article helpful? Yes No 0 out found this helpful

Shall we send you a message when we have discounts available?

Remind me later

Thank you! Please check your email inbox to confirm.

Oops! Notifications are disabled.