Determine the necessary mass, volume, or concentration for preparing a solution.
BioReagent BioReagent para flujos de trabajo cromatográficos y analíticos sensibles que requieren una interferencia mínima de la línea base.
Store at 2-8°C,Protected from light Envío Wet ice. Consulte el COA específico del lote para conocer las especificaciones exactas.
SDS, COA, ficha técnica y hoja de especificaciones disponibles para descargar. COA específico del lote accesible mediante búsqueda por número de lote.
Citado en 0 publicaciones revisadas por pares sobre cromatografía, síntesis orgánica y reacciones de acoplamiento cruzado.
In acidic solution, under the action of decomposing agent and high temperature, inorganic phosphate and organic phosphate are hydrolyzed into orthophosphate. Orthophosphate reacts with ammonium molybdate to form phosphomolybdic acid, which is then reduced to phosphomolybdenum blue in the presence of a reducing agent, with a characteristic absorbance peak at 710 nm.
|
Notes: Before formal detection, it is recommended to conduct preliminary trials with 2–3 samples with expected large differences in phosphate content.
Self-Prepared Consumables and Instruments
1.Microplate reader or visible spectrophotometer (capable of absorbance detection at 710 nm)
2.96-well microplate or micro glass cuvettes, adjustable pipettes and pipette tips
3.Tabletop centrifuge, 30–50 mesh sieve, thermostatted water bath
4.Deionized water
Reagent Preparation
Reagent I: Ready-to-use; equilibrate to room temperature before use; store at 2-8 °C.
Reagent II: Ready-to-use; equilibrate to room temperature before use; store at 2-8 °C.
Reagent III: Ready-to-use; equilibrate to room temperature before use; store at 2-8 °C protected from light.
Working Reagent IV: Prepare immediately before use. Add 5.6 mL deionized water and dissolve completely, then add Reagent V and mix thoroughly. Working Reagent IV contains a small amount of insoluble precipitate; shake well before use. Stable for 1 week when stored at 2-8 °C.
Reagent V: Ready-to-use; equilibrate to room temperature before use; store at 2-8 °C.
Notes: Reagent I, Reagent II, Reagent III and Reagent V are irritant. Adequate personal protective equipment must be worn during all operations.
Standard Solution: Ready-to-use; equilibrate to room temperature before use; store at 2-8 °C.
Standard Solution Preparation: Dilute the 1 μmol/mL standard stock solution stepwise as shown in the table below.
|
Notes: A standard curve must be generated alongside each independent experiment. Diluted standard solutions are unstable and must be used within 4 hours of preparation.
Sample Preparation
Fresh samples are recommended for immediate detection. If testing cannot be performed promptly, samples can be stored at −80 °C for several weeks. Control thawing temperature and duration strictly; complete thawing within 4 hours at room temperature.
1.Air-dry fresh soil samples and sieve through a 30–50 mesh screen.
Mix soil mass (g) : deionized water volume (mL) : Reagent I volume (mL) : Reagent II volume (mL) = 1:10:1:2.
Practical operation: Weigh approximately 0.1 g soil sample, add 1 mL deionized water, followed by 100 μL Reagent I and 200 μL Reagent II. Shake in a 95 °C boiling water bath for 30 min, centrifuge at 10,000 g, 25 °C for 10 min, collect supernatant for subsequent detection.
Assay Procedure
1.Preheat the microplate reader or visible spectrophotometer for ≥30 min, set detection wavelength to 710 nm. Zero the spectrophotometer with deionized water if applicable.
2.Assay layout (operations performed in 96-well microplate or micro glass cuvettes):
|
3. Mix thoroughly and incubate at 25 °C for 10 min. Measure absorbance at 710 nm and record the absorbance values as Ablank, Astandard and Asample respectively.
Result Calculation
Notes: Two equivalent formulas are provided, namely the full derivation formula and the simplified formula.

Figure 1. Total phosphate content in soil detected by this kit
Disclaimer
This product is for research purposes only and is not suitable for clinical diagnosis. Please wear a lab coat and disposable gloves during operation to ensure your safety and health.
| Item No. | Appearance | Components | 48T | 96T | Storage |
| S1522127A | Liquid | Reagent Ⅰ | 10 mL | 20 mL | 2-8 ℃ |
| S1522127B | Liquid | Reagent Ⅱ | 15 mL | 30 mL | 2-8 ℃ |
| S1522127C | Liquid | Reagent Ⅲ | 2.8 mL | 5.6 mL | 2-8 ℃. Store in the dark. |
| S1522127D | Powder | Reagent Ⅳ | 1 vial | 1 vial | 2-8 ℃ |
| S1522127E | Liquid | Reagent V | 44.8 µL | 44.8 µL | 2-8 ℃ |
| S1522127F | Liquid | Standard | 1 mL | 1 mL | 2-8 ℃ |
Documento completo sobre peligros, manipulación, almacenamiento y cumplimiento normativo.
<<<<<<< HEAD Descargar SDS → ======= Descargar SDS → >>>>>>> release248-p3Datos de calidad específicos del lote. Introduzca su número de lote para obtener el COA exacto.
Buscar COA →Resumen de referencia rápida de las especificaciones y aplicaciones del producto.
Ver ficha técnica →Atributos de calidad completos y criterios de aceptación para esta calidad.
Ver hoja de especificaciones →Nuestra guía de selección de calidades cubre la pureza, el estado del estabilizador y la idoneidad de la aplicación para todas las variantes de nuestro catálogo.
Ver la guía de calidades BioReagent →