Experiments in methodological evaluation

Summary

I. Intra-lot reproducibility test

Purpose: To examine the random error of the experimental method and evaluate the precision of the method.

II. Recovery test

Purpose: To detect the proportional systematic error to measure the accuracy of the analytical method .

III. Interference test

Purpose: to examine the constant systematic error of the method to evaluate the accuracy of the candidate method

This experiment is derived from the experiment guide of Mudanjiang Medical College, undergraduate 5-year testing program.

Operation method

Experiments in methodological evaluation

Move

I. Intra-lot reproducibility test

1. Repeat the determination of the same sample 20 times within a short period of time in accordance with the prescribed method.

2. Statistical processing: the results of the measurement. Calculate the mean value (X), standard deviation (SD), coefficient of variation (CV) according to the following formula.

X = ∑X/n SD = ∑(X-X)2/n-1 CV = SD/Y × 100%

3. Evaluation: general analysis CV value less than 50%, precision analysis CV value less than 2

Recovery test

1. Sample preparation:

Base sample: serum 2.0ml + distilled water 0.1ml

Low concentration recovery sample: serum 2.0ml + 4mmol/L calcium standard 0.1ml

High concentration recovery sample: serum 2.0m1 + 25mmo1/L calcium standard solution 0.1m1

2. Calculate the added concentration according to the following formula

Addition concentration = standard solution concentration × standard solution volume / serum volume + standard solution volume

3. Measure the concentration in each sample tube more than 3 times.

4. Calculation. Calculate the sample recovery concentration and recovery rate separately

5. Recovered Concentration = Measured Concentration of Recovered Sample - Measured Concentration of Base Sample

Recovery rate = Recovered concentration / added concentration x 100%.

III. Interference test

1. Sample preparation:

Base sample: blood please 1.8m1 + 0.2ml

Low concentration recovery sample: serum 1.8m1 + 63umol/L creatinine standard 0.2ml

High concentration recovery: serum 1.8m1 + 414umol/L creatinine standard 0.2ml

2. Calculate the added concentration according to the following formula

Addition concentration = standard solution concentration × standard solution volume / serum volume + standard solution volume

3. Calculate the concentration in each sample tube

4. Calculate the interference value:

Interference value = measured concentration of interfering sample - measured concentration of interfering sample

5. How to eliminate the interference (discussion)


For more product details, please visit Aladdin Scientific website.

https://www.aladdinsci.com/

Categories: Protocols
Explore topics: Clinical trial

Da — when not otherwise indicated, molecular weight units are daltons.   Mw — weight-average molecular weight.   Mn — number-average molecular weight.

Products are supplied for research and development use only. Not for use in humans, animals, diagnosis, or therapy.

Cite this article

Aladdin Scientific. "Experiments in methodological evaluation" Aladdin Knowledge Base, updated Dec 24, 2024. https://www.aladdinsci.com/us_en/faqs/experiments-in-methodological-evaluation-en.html
Was this article helpful? Yes No 0 out found this helpful

Shall we send you a message when we have discounts available?

Remind me later

Thank you! Please check your email inbox to confirm.

Oops! Notifications are disabled.