Item-specific, tested applications and dilutions: Not specified for this item; refer to CoA/Spec Sheet.
General protocols (literature/general; adapt to certificate instructions)
DLS calibration check
Equilibrate the CRM to measurement temperature (e.g., 25°C). Gently invert; avoid bubbles.
Prepare a dilution in 0.1–0.2 µm filtered DI water to reach the manufacturer’s recommended count rate/attenuation. Mix by gentle inversion.
Rinse a clean, dust-free cuvette three times with the dilution. Fill and cap. Allow 5–10 min to equilibrate. Measure ≥3 runs; assess z-average and PDI against certified values.
Laser diffraction verification
Select optical model using the CRM’s refractive index. Circulate dispersant to achieve baseline. Add drops of CRM until obscuration target is reached.
Acquire triplicate measurements at specified pump/stir speeds. Compare D10/D50/D90 or volume mean to certified size.
NTA size/count check
Dilute to reach recommended particle/frame. Record multiple videos at consistent camera settings. Analyze and compare mean/modal size to certificate.
Good practice
Record lot number, expiry, dilution recipe, temperature, and instrument settings with each run.
Use fresh aliquots; avoid returning used material to the stock to prevent contamination.
Biological Roles
Not typically applicable to this product.
Particle size CRMs are metrological standards rather than biomolecules. They do not possess inherent biological roles or metabolic functions.
If used in bioanalytical method qualification (e.g., sizing extracellular vesicles), their role is as a physical size standard, not as a biological reagent.
Buffer Applications
Limited applicability.
These materials are not buffering agents. However, dispersion and dilution are often performed in simple media (e.g., filtered deionized water) or specified low-ionic-strength buffers to maintain colloidal stability.
For this item, recommended diluents, ionic strength, and pH windows are not specified; refer to CoA/Spec Sheet.
General tips (literature/general)
Use freshly filtered (0.1–0.2 µm) water to minimize dust that can skew light-scattering results.
Avoid high-salt buffers unless the CRM certificate endorses them; electrolyte screening can induce aggregation or alter hydrodynamic size.
Match dispersant viscosity and temperature to the conditions under which certification was performed for best agreement.
Green Alternatives
Context
Green chemistry considerations are not central for particle size CRMs, but selection of medium and handling can reduce environmental impact.
General guidance (literature/general)
Prefer aqueous dispersions when possible to avoid volatile organic compounds (VOCs) used in some nonaqueous standards.
Minimize waste by preparing only the volumes needed for a session and by using closed cuvettes to extend working life of dilutions.
Reuse policy: Do not return used dilutions to stock to avoid contamination; instead, plan small-volume dilutions to reduce disposal.
Disposal: Follow institutional and local regulations. Many latex/silica dispersions can be coagulated/flocculated and filtered for solid waste disposal; avoid drain disposal unless permitted.
Comparison (general)
Aqueous medium: Low toxicity, minimal VOCs, simple waste streams; may require biocides which introduce additional EHS considerations.
Alcohol-based medium (IPA/EtOH): Faster drying and better wetting for imaging but flammable and VOC-emitting; manage with closed containers and fume hoods.
Item-specific note
The exact medium and preservatives for this item are not specified; consult the CoA/Spec Sheet and SDS for environmentally relevant constituents before planning waste management.
Pharmaceutical Uses
Formulation/Manufacturing context (no clinical claims)
Particle size CRMs may be used in pharmaceutical quality control laboratories to qualify and calibrate sizing instruments used for raw materials, suspensions, inhalation powders, and injectable dispersions.
They can support method validation parameters (accuracy, precision, range) for compendial or in-house procedures assessing particle size distributions.
Item-specific note
Pharmacopeial status, compliance, or specific suitability claims for this item are not specified; refer to CoA/Spec Sheet and regulatory guidelines.
Good practices (literature/general)
Maintain traceability: Record CRM lot numbers in validation protocols and OOS investigations.
Evaluate matrix effects: Demonstrate that measurements in product-relevant media are equivalent to CRM measurements or account for differences.
Establish requalification intervals for instruments using CRMs bracketing the routine measurement range.
Physical Properties
Item-specific facts (from Product Data)
Physical state/appearance: Not specified for this item; refer to CoA/Spec Sheet.
Density, refractive index, zeta potential, dispersant, solids content: Not specified for this item; refer to CoA/Spec Sheet.
Literature/general values (context only; not item-specific)
Polystyrene latex microspheres: density ~1.05 g/cm³; refractive index ~1.59 @ 589 nm (literature).
Amorphous silica microspheres: density ~2.0–2.2 g/cm³; refractive index ~1.45 @ 589 nm (literature).
Typical certified size ranges available commercially span nanometers to tens/hundreds of micrometers (e.g., ~20 nm to ~100 µm), with CV often <5% for monodisperse lots (literature).
Dispersibility: Often water or alcohols for latex; silica may require aqueous buffers with pH control to minimize aggregation (literature).
Notes for users
Optical particle-sizing requires approximate refractive index of both particles and medium; use supplier-provided n and k values when modeling Mie scattering. For this item, consult CoA/Spec Sheet for optical constants.
Temperature and ionic strength can affect hydrodynamic size in DLS; verify measurement temperature and viscosity of the dispersant.
Quality and Grades
Item-specific details (from Product Data)
Grade/Purity: Not specified for this item; refer to CoA/Spec Sheet.
What “Certified Reference Material (CRM)” typically entails (literature/general)
Metrological framework: CRMs are characterized by a technically valid procedure, certified for one or more specified properties (e.g., particle size), with an associated uncertainty and statement of metrological traceability.
Accreditation: High-quality particle size CRMs are often produced under ISO 17034 (reference material producers) and with measurements under ISO/IEC 17025 (testing/calibration labs). Accreditation status for this item is not specified; refer to documentation.
Certificate contents: Certified mean/modal size, expanded uncertainty (k-factor), homogeneity and stability assessment, measurement methods (e.g., TEM/SEM, AFM, SEM image analysis, DLS, laser diffraction), traceability route, and instructions for use.
Fit-for-purpose: Select a CRM close to your operational size range and optical model. For laser diffraction, match refractive index; for DLS, match hydrodynamic regime and polydispersity.
Stabilizers/Preservatives: Some aqueous standards include surfactants/biocides to prevent aggregation or microbial growth. Presence/identity for this item is not specified; refer to CoA/Spec Sheet.
Practical notes
Use unopened aliquots for QC trending to minimize use uncertainty.
Record environmental conditions (temperature, viscosity, ionic strength) when measuring to maintain comparability with the certified state.
Reaction and Applications
Applicability to this product
These materials are standards, not reagents for chemical synthesis; no “reactions” apply. Instead, applications focus on instrument qualification, calibration, and method validation.
Key applications (literature/general)
Dynamic Light Scattering (DLS): Verify size accuracy, repeatability, and resolution across submicron ranges; assess instrument alignment and cuvette cleanliness.
Laser Diffraction (LD): Validate size distributions from ~0.1–1000 µm; test alignment, obscuration settings, and optical model robustness.
Nanoparticle Tracking Analysis (NTA): Check sizing and concentration linearity; verify camera focus and thresholding.
Coulter/Resistive Pulse Sensing: Validate aperture selection and coincidence corrections.
Method transfer and QC trending: Use the same CRM lot to benchmark inter-instrument variability and operator effects.
Practical tips
Homogenization: Gently invert or roll; brief low-power sonication if permitted in the CoA. Avoid frothing.
Cleanliness: Use low-particulate labware; rinse cuvettes/flow cells with filtered medium.
Concentration: Work within the certificate’s recommended attenuator/obscuration or count rate targets to avoid multiple scattering or coincidence.
Documentation: Record lot number, expiry, storage history, and dilution recipe to maintain traceability.
Reaction Conditions
Not applicable to chemical reactions.
General usage conditions for sizing (literature/general)
Temperature control: DLS and NTA results are temperature-dependent via viscosity; stabilize at 20–25°C or as specified in the certificate.
Concentration: Aim for instrument-recommended count rates/obscuration (e.g., single scattering regime for DLS; low coincidence for Coulter-type methods). Exact targets vary by instrument.
Sonication: If allowed, brief, low-power sonication helps break weak agglomerates; avoid over-sonication that can alter particle integrity.
Filtration: Use syringe filters (0.2–0.45 µm) only if the certified size is much smaller than the pore size and if permitted by the certificate; filtration can bias distributions.
Item-specific parameters such as certified size, acceptable media, and use instructions are not specified for this item; refer to CoA/Spec Sheet.
Safety and Handling
Item-specific hazards (from Product Data)
GHS signal word, hazard statements, classification, pictograms: Not specified for this item; refer to SDS.
Storage conditions: Room temperature (per Product Data).
Shipped: Normal conditions.
General safety guidance (literature/general)
Handle CRMs as laboratory chemicals. Avoid generating dust/aerosols. For powders, use local exhaust; for dispersions, prevent inhalation of mists.
PPE: Safety glasses, lab coat, disposable nitrile gloves. Use splash protection when preparing dilutions or sonication.
Incompatibilities: Dispersions can be destabilized by high ionic strength, surfactant mismatch, extreme pH, and organic solvent blends not recommended by the supplier. Avoid cationic surfactants with negatively charged latex unless specified.
First aid (overview; defer to SDS): Eye contact—rinse cautiously with water for several minutes. Skin—wash with soap and water. Inhalation—move to fresh air. Ingestion—rinse mouth; seek medical advice if symptoms persist.
Spill response: For powders, avoid dry sweeping; dampen and collect. For liquids, contain and absorb on inert material; prevent release to drains until assessed.
Special risks: Ultrasonic bath/probe can aerosolize or heat dispersions; use covers and control power settings. When drying droplets for microscopy, avoid hot plates that can char organics; use low heat.
Always consult the product’s SDS and CoA/Spec Sheet for definitive hazard and handling instructions.
Solvent Selection
Applicability to this product
Solvent selection is not typically the primary consideration for particle size CRMs; they are often supplied as ready-to-use aqueous dispersions or as dry powders intended for dispersion in specified media. For this item, the supplied medium and any recommended diluents are not specified; refer to CoA/Spec Sheet.
General guidance (literature/general)
Aqueous dispersions: Use deionized, filtered water (0.1–0.2 µm) with low ionic strength to minimize DLVO-driven aggregation. Adjust pH within the stability window stated in the CoA.
Organic-compatible standards: Some CRMs are validated in alcohols (e.g., ethanol/IPA) or specific nonaqueous media for LD or imaging; match refractive index models accordingly.
Additives: Only add surfactants (e.g., 0.01–0.05% nonionic) if recommended by the certificate; unintended additives can alter hydrodynamic size and zeta potential.
Comparison (general)
Water: Highest instrument compatibility for DLS/NTA; low toxicity; sensitive to CO2 absorption and ionic contaminants.
IPA/Ethanol: Faster drying for microscopy grids; lower surface tension aids wetting; flammable and can change particle–particle interactions.
Buffered saline: Improves biological relevance but increases screening; may induce aggregation unless the CRM is rated for it.
Storage and Reconstitution
Item-specific storage and shipping (from Product Data)
Storage conditions: Room temperature.
Shipped in: Normal conditions.
Item-specific details not provided
Appearance, dispersant, solids content, preservative/biocide, shelf life/expiry, and reconstitution instructions: Not specified for this item; refer to CoA/Spec Sheet.
General guidance (literature/general)
Storage
Keep containers tightly closed to limit evaporation/contamination. Store upright and protect from freezing if the medium is aqueous (unless the certificate permits freezing).
Avoid prolonged exposure to light/UV that may degrade polymeric matrices or preservatives.
Do not store working dilutions for extended periods unless validated; microbial growth and evaporation change concentration and size distribution.
Reconstitution/Preparation
For dry powders, if applicable: Allow to equilibrate to room temperature before opening to minimize moisture condensation. Disperse in specified medium, using gentle stirring and, if permitted, brief sonication.
For dispersions: Gently invert/roll to homogenize. Do not vortex aggressively to avoid foaming and air entrapment.
Traceability
Record opening date and maintain lot-specific logs. Use within the validity period indicated on the CoA when available.
CAS, CID, InChIKey, SMILES, Molecular formula, Molecular weight: Not specified for this item; refer to CoA/Spec Sheet.
General description (literature/general)
These materials are typically monodisperse particulate standards used to verify and calibrate particle-sizing instruments (e.g., laser diffraction, DLS, NTA, image analysis, electron microscopy).
Common matrices: polymeric latex (e.g., polystyrene), silica, or metal-oxide spheres; may be supplied as aqueous dispersions or dry powders. Specific composition for this item is not specified.
Structural features: Spherical morphology (most common for size standards), narrow size distribution with a certified mean diameter and uncertainty; some products include multimodal blends for resolution checks.
Identity attributes usually provided in a CRM’s documentation: certified modal/mean particle size, distribution width, uncertainty budget, measurement method(s), and metrological traceability. For this item, consult CoA/Spec Sheet for exact identity attributes.
Synthetic Utility
Not applicable.
Particle size CRMs are standards rather than synthetic building blocks or reagents. They are not typically used to perform chemical transformations.
For research involving nanoparticle synthesis, CRMs may serve as independent controls to verify sizing methods, but they are not reactants or catalysts.
Target Specificity
Not applicable.
This product is not an antibody, probe, or affinity reagent and does not have biological target specificity.
Wir verwenden Cookies, um die ordnungsgemäße Funktion der Website sicherzustellen und, soweit zulässig, Ihr Nutzererlebnis zu verbessern. Sie können Ihre Einstellungen jederzeit unter „Einstellungen“ verwalten. Weitere Informationen finden Sie in unserer Cookie-Richtlinie.
Dürfen wir Sie über Rabattaktionen informieren?
Später erinnern
Vielen Dank! Bitte überprüfen Sie Ihren E-Mail-Posteingang, um zu bestätigen.
Products are supplied to verified businesses, institutions, and qualified professionals for research and development use only. Not for use in humans, animals, diagnosis, or therapy.