Protocols

Factor VII Related Antigen Content Assay

Summary

The turbidity of the immunocomplexes between coagulation factor VIII-related antigen (ⅧR; Ag) and its corresponding specific antibody is determined by the transmission method, and the turbidity is proportional to the concentration of ⅧR:Ag in the serum. This experiment is from the laboratory instruction of Mudanjiang Medical College undergraduate 5-year testing program.

Operation method

Factor VII-related antigen content assay

Principle

The turbidity of the immunocomplexes between coagulation factor VIII-related antigen (VIIIR; Ag) and its corresponding specific antibody is determined by transmission assay and is proportional to the serum concentration of IV VIIIR:Ag.

Move

Experimental method: manual, semi-automatic measurement


Mixing 37 ℃ water bath for 20 minutes, 340 nm with antiserum zero point determination of each tube A value D

Data processing: ⅧR: Ag content = (A-value of measurement tube / A-value of reference tube)×100%.

Normal value: 60-150%.

Caveat

1. When the turbidity of the specimen is too high, the dilution ratio of the specimen should be regulated and re-specified.

2. high blood fat. Jaundice and mild hemolysis have little effect on the results of this method.

3. Linear range of the method: 0-200%.Clinical significance: ⅧR: Ag content increased: seen in coronary heart disease myocardial infarction, cerebral thrombosis kidney disease, etc.. Decrease in ⅧR:Ag level: seen in vascular pseudohaemophilia, haemophilia carriers, etc.


For more product details, please visit Aladdin Scientific website.

https://www.aladdinsci.com/

Categories: Protocols
Explore topics: Clinical trial

Da — when not otherwise indicated, molecular weight units are daltons.   Mw — weight-average molecular weight.   Mn — number-average molecular weight.

Products are supplied for research and development use only. Not for use in humans, animals, diagnosis, or therapy.

Cite this article

Aladdin Scientific. "Factor VII Related Antigen Content Assay" Aladdin Knowledge Base, updated Dec 24, 2024. https://www.aladdinsci.com/us_en/faqs/factor-vii-related-antigen-content-assay-en.html
Was this article helpful? Yes No 0 out found this helpful

Shall we send you a message when we have discounts available?

Remind me later

Thank you! Please check your email inbox to confirm.

Oops! Notifications are disabled.