Recombinant CD42b Antibody (FITC) Flow Cytometry Protocol
Recombinant CD42b Antibody (FITC) Flow Cytometry Protocol
Recombinant CD42b Antibody (FITC) is a fluorescein isothiocyanate-conjugated recombinant antibody targeting the CD42b (GPIbα) platelet surface glycoprotein, supplied at 0.5 µg per test for direct immunofluorescence staining and flow cytometric analysis.
- Product Name
- Recombinant CD42b Antibody (FITC)
- SKU
- Ab229534
- Amount per Test
- 0.5 µg per test
- Fluorochrome
- FITC (Fluorescein Isothiocyanate)
- Antibody Format
- Recombinant, directly conjugated
- Unlabeled Parent Clone
- Ab222899Z — Recombinant CD42b Antibody
- Recommended Application
- Flow Cytometry (direct staining)
Overview of CD42b and the FITC Conjugate
CD42b, also known as glycoprotein Ibα (GPIbα), is a type I transmembrane glycoprotein constitutively expressed on the surface of platelets and megakaryocytes. It forms a non-covalent complex with CD42c (GPIbβ), CD42a (GPIX), and CD42d (GPV) to constitute the GPIb-IX-V complex, which serves as the primary receptor for von Willebrand factor (vWF) and mediates platelet adhesion to damaged vascular subendothelium under high shear conditions. CD42b is therefore a widely used marker for platelet identification, activation studies, and clinical research into bleeding disorders such as Bernard-Soulier syndrome.
Recombinant CD42b Antibody (FITC) (Ab229534) is produced from a recombinant expression system using the sequence of the unlabeled parent antibody Ab222899Z and is directly conjugated to FITC for single-step, no-secondary-antibody staining. Each vial is formatted as a per-test reagent containing 0.5 µg of conjugated antibody, enabling reproducible direct immunofluorescence staining suitable for flow cytometric analysis.
Materials Required
Before beginning the staining protocol, ensure the following materials are available:
- Recombinant CD42b Antibody (FITC) (Ab229534), 0.5 µg per test
- Phosphate-buffered saline (PBS), pH 7.2–7.4
- Flow cytometry staining buffer (e.g., PBS supplemented with 1–2% BSA or 2% FBS)
- Fixation solution, if intracellular co-staining or fixed-cell analysis is required (e.g., 1–4% paraformaldehyde in PBS)
- Microcentrifuge tubes or V-bottom 96-well plate
- Centrifuge capable of 300–400 × g
- Flow cytometer equipped with a 488 nm excitation laser and a 530/30 nm (or equivalent) emission filter for FITC detection
- Isotype control antibody (FITC-conjugated, matched isotype) for gating reference
All reagents should be at room temperature before use unless otherwise specified. Protect FITC-conjugated reagents from prolonged light exposure to prevent photobleaching.
Step-by-Step Staining Protocol
The following protocol is intended for direct surface staining of platelets or platelet-containing peripheral blood samples. Adjust cell numbers and volumes as required for your specific experimental system.
- Sample preparation: Collect whole blood or isolate platelet-rich plasma (PRP) according to standard laboratory procedures. If using whole blood, transfer 50–100 µL into a staining tube. If using isolated cells, adjust to approximately 1 × 106 cells per test in 100 µL of staining buffer.
- Blocking (optional): To reduce non-specific binding, incubate cells with an Fc receptor blocking reagent for 10 minutes at 4 °C or on ice. This step is especially recommended for samples containing monocytes or other Fc receptor-expressing cells.
- Antibody addition: Add the full content of one test (0.5 µg) of Recombinant CD42b Antibody (FITC) (Ab229534) directly to the cell suspension. Mix gently by pipetting.
- Incubation: Incubate for 20–30 minutes at 4 °C or room temperature, protected from light. Do not exceed 30 minutes to minimize non-specific signal.
- Washing: Add 2 mL of staining buffer and centrifuge at 300–400 × g for 5 minutes. Carefully aspirate the supernatant without disturbing the cell pellet. Repeat the wash step once.
- Resuspension: Resuspend the cell pellet in 200–500 µL of staining buffer (or fixation solution if fixed analysis is required). Keep samples on ice and protected from light until acquisition.
- Data acquisition: Acquire samples on a flow cytometer with a 488 nm excitation laser. Detect FITC fluorescence in the FL1 channel (530/30 nm filter or equivalent). Record a minimum of 10,000 events within the gated population of interest.
- Gating: Use forward scatter (FSC) and side scatter (SSC) parameters to identify the platelet population (small, low-granularity events). Set gates using the isotype control or fluorescence-minus-one (FMO) control to distinguish positive from negative populations.
Include appropriate controls in every experiment: unstained cells, isotype control (FITC-conjugated), and, where applicable, a known positive control sample.
Troubleshooting and Tips
The following guidance addresses common issues encountered when using FITC-conjugated antibodies in flow cytometry:
- High background / non-specific staining: Ensure the Fc receptor blocking step is performed. Increase the number of wash steps or use a more stringent wash buffer. Confirm that the instrument voltage settings are appropriate and that compensation is correctly applied for multi-color panels.
- Low or absent signal: Verify that the FITC channel is correctly configured (488 nm excitation, ~530/30 nm emission). Check sample viability; dead cells exhibit high autofluorescence and non-specific antibody binding. Ensure the antibody was stored correctly at 2–8 °C, protected from light, and was not subjected to freeze-thaw cycles.
- Photobleaching: FITC is susceptible to photobleaching. Keep all FITC-stained samples and the antibody stock protected from direct light at all stages of the protocol.
- Platelet activation during processing: Blood collection and processing can activate platelets and alter surface receptor expression. Use gentle handling, avoid vortexing, and process samples promptly after collection. Consider adding an activation inhibitor (e.g., prostaglandin E1 or theophylline) to collection tubes if platelet activation is a concern.
- Per-test format: Each test of Ab229534 contains 0.5 µg of Recombinant CD42b Antibody (FITC). Do not pool multiple tests unless titration experiments confirm that higher concentrations are appropriate for your cell type and density.
Storage, Handling, and Quality Notes
Storage: Store Recombinant CD42b Antibody (FITC) (Ab229534) at 2–8 °C. Protect from light at all times. Do not freeze the product, as freeze-thaw cycles can denature the antibody and cause aggregation that reduces performance and may clog instrument fluidics.
Handling: Allow the vial to equilibrate to room temperature briefly before opening. Gently mix by inversion or flicking; do not vortex. Use aseptic technique to avoid microbial contamination of the stock.
Shelf life: Refer to the expiry date printed on the product label. Performance beyond the stated expiry date cannot be guaranteed.
Recombinant format advantage: Because Ab229534 is derived from a recombinant expression system (parent clone Ab222899Z), it offers lot-to-lot consistency superior to that of hybridoma-derived antibodies, minimizing the need for re-optimization between lots.
Quality control: Each lot should be tested by the end-user with appropriate positive and negative controls before use in critical experiments. If performance deviates from expectations, contact Aladdin Scientific technical support and retain the product for evaluation.
Frequently asked questions
What is the recommended amount of Recombinant CD42b Antibody (FITC) to use per test?
Each test of Ab229534 contains 0.5 µg of Recombinant CD42b Antibody (FITC), which represents the manufacturer-defined amount for a single staining reaction under standard conditions. Users should perform titration experiments to confirm optimal concentration for their specific cell type, sample matrix, and instrument settings.
Can this antibody be used for applications other than flow cytometry?
Ab229534 is formatted and validated for flow cytometric analysis using direct FITC fluorescence detection. FITC conjugation is generally not compatible with applications such as Western blotting or IHC on paraffin-embedded tissue due to signal loss during fixation and processing. Suitability for other applications should be independently verified by the end-user; consult Aladdin Scientific technical support for guidance.
What controls should be included when using this antibody?
At minimum, include an unstained cell control, an isotype-matched FITC-conjugated control antibody to assess non-specific background, and a known CD42b-positive sample (e.g., human platelets) as a positive control. Fluorescence-minus-one (FMO) controls are also recommended in multi-color panels to accurately define gating boundaries for the FITC channel.
What is the relationship between Ab229534 and Ab222899Z?
Ab229534 (Recombinant CD42b Antibody, FITC) is the FITC-conjugated derivative of Ab222899Z (Recombinant CD42b Antibody, unlabeled). The recombinant antibody sequence is identical; the only difference is the covalent attachment of the FITC fluorochrome to Ab229534, enabling direct detection without a secondary antibody step.
How should I store the antibody to maintain optimal performance?
Store Ab229534 at 2–8 °C, protected from light at all times. Do not freeze. Avoid repeated opening of the vial to minimize contamination risk. Discard any unused portion after the expiry date printed on the label.
References
- López JA, Andrews RK, Afshar-Kharghan V, Berndt MC. Bernard-Soulier syndrome. Blood. 1998;91(12):4397-4418. PMID:9616133
- Ruggeri ZM. Von Willebrand factor, platelets and endothelial cell interactions. J Thromb Haemost. 2003;1(7):1335-1342. PMID:12871276
- Bhatt DL, Topol EJ. Scientific and therapeutic advances in antiplatelet therapy. Nat Rev Drug Discov. 2003;2(1):15-28. PMID:12509759
Products are supplied for research and development use only. Not for use in humans, animals, diagnosis, or therapy.