Determine the necessary mass, volume, or concentration for preparing a solution.
BioReagent, for enzyme immunoassay(ELISA) BioReagent,for Enzyme immunoassay(ELISA) for sensitive chromatographic and analytical workflows requiring minimal baseline interference.
Store at 2-8°C Ships Wet ice Check lot-specific COA for exact specifications.
SDS, COA, datasheet, and spec sheet available for download. Lot-specific COA accessible via lot number lookup.
Cited in 0 peer-reviewed publications across chromatography, organic synthesis, and cross-coupling reactions.
Lecanemab is a monoclonal antibody drug indicated for the treatment of Alzheimer’s disease. It is an anti-amyloid-β antibody administered intravenously to patients with mild cognitive impairment or mild dementia. In clinical trials, it demonstrated moderate efficacy in slowing the relative rate of cognitive decline compared with placebo. As a humanized monoclonal antibody derived from murine parent antibody mAb158, Lecanemab recognizes amyloid-β protofibrils and inhibits amyloid-β deposition in preclinical animal models of Alzheimer’s disease.
This kit adopts the indirect enzyme-linked immunosorbent assay (ELISA) method. Microwells pre-coated with human amyloid-beta antigen are sequentially loaded with samples, standards, and HRP-conjugated secondary antibody, with incubation and washing steps in between. Color development is performed using TMB substrate: TMB turns blue upon catalysis by horseradish peroxidase (HRP), and shifts to final yellow after acid termination. The color intensity is positively correlated with the concentration of Lecanemab in the sample. The absorbance (OD value) is measured at 450 nm using a microplate reader to calculate sample concentration.
1. Strictly follow the specified incubation time and temperature to guarantee accurate results. All reagents must be equilibrated to room temperature (20–25°C) before use. Refrigerate reagents immediately after use.
2. Improper plate washing may lead to inaccurate readings. Ensure complete removal of residual liquid in wells prior to substrate addition. Avoid prolonged drying of microwells throughout the procedure.
3. Wipe residual liquid and fingerprints from the bottom of the microplate, as contaminants will interfere with OD values.
4. The TMB substrate solution shall remain colorless; discard substrate solution that has turned blue.
5. Prevent cross-contamination between reagents and samples to avoid erroneous test results.
6. Avoid direct intense light exposure during reagent storage and incubation.
7. All reaction reagents must not come into contact with bleach solvents or their volatile fumes; bleach components will destroy the biological activity of kit reagents.
8. Do not use expired products; components from different lot numbers or batch codes cannot be mixed.
9. Recombinant proteins from external sources may fail to be recognized due to mismatched antibody affinity with kit capture antibodies.
10. Handle all potentially infectious samples properly, and dispose of specimens and testing devices in accordance with standardized biosafety protocols.
1. The kit’s detectable concentration range does not equal the analyte concentration range in raw samples. It is recommended to estimate sample concentrations via published literature and conduct pre-tests to confirm actual analyte levels. Dilute or concentrate samples appropriately if analyte concentrations fall outside the linear range.
2. If your sample matrix is not listed in this manual, perform a pre-test to verify assay compatibility.
3. Serum: Collect whole blood in serum separation tubes, incubate at room temperature for 2 hours or store overnight at 2–8°C, then centrifuge at 1000×g for 20 minutes. Collect the supernatant for immediate testing, or aliquot and store at -20°C / -80°C; avoid repeated freeze-thaw cycles.
4. Plasma: Collect blood using EDTA or heparin anticoagulant. Centrifuge specimens within 30 minutes post-collection at 1000×g, 2–8°C for 15 minutes. Harvest supernatant for immediate testing, or store at -20°C / -80°C; avoid repeated freeze-thaw cycles.
5. Tissue Homogenate: Rinse tissue with pre-cooled PBS (0.01 M, pH=7.4) to remove residual blood (lysed erythrocytes in homogenate interfere with detection). Weigh and mince tissue, then mix tissue with PBS at a standard weight-to-volume ratio of 1:9 (e.g., 1 g tissue in 9 mL PBS; adjust volume as needed and record details). Protease inhibitor cocktail is recommended for PBS. Transfer the mixture to a glass homogenizer and grind thoroughly on ice, or use a mechanical homogenizer. For complete cell lysis, sonicate the homogenate or perform repeated freeze-thaw cycles. Centrifuge the homogenate at 5000×g for 5–10 minutes and collect supernatant for testing.
6. Cell Culture Supernatant: Centrifuge culture media at 1000×g for 20 minutes, collect supernatant for immediate testing, or store at -20°C / -80°C; avoid repeated freeze-thaw cycles.
7. Cell Lysate: For adherent cells, wash gently with pre-cooled PBS, digest with trypsin, then harvest cells via centrifugation at 1000×g for 5 minutes. Suspension cells can be pelleted directly by centrifugation. Wash harvested cells 3 times with pre-cooled PBS, resuspend 1×10⁶ cells in 150–200 μL PBS (protease inhibitors recommended; reduce PBS volume for low-abundance analytes). Lyse cells via repeated freeze-thaw or sonication. Centrifuge lysate at 1500×g, 2–8°C for 10 minutes and retain supernatant.
8. Other Biological Matrices: Centrifuge samples at 1000×g for 20 minutes and collect supernatant for testing.
9. Sample Appearance: Samples must be clear and transparent; all suspended particulates must be removed by centrifugation.
10. Sample Storage: Samples to be tested within 1 week can be stored at 4°C. For delayed testing, aliquot samples into single-use volumes and freeze at -20°C (stable for 1 month) or -80°C (stable for 6 months). Avoid repeated freeze-thaw cycles. Hemolyzed samples are unsuitable for this assay, as hemolysis distorts final readings.
Estimate analyte concentration in advance. Follow the dilution schemes below if sample dilution is required:
100-fold dilution (single step): Mix 5 μL sample with 495 μL Assay Diluent.
1000-fold dilution (two-step): Step 1: Dilute 5 μL sample into 95 μL diluent (20×). Step 2: Transfer 5 μL of the 20× diluted sample into 245 μL diluent (50×), achieving total 1000-fold dilution.
100000-fold dilution (three-step): Step 1: Mix 5 μL sample with 195 μL diluent (40×). Step 2: Transfer 5 μL 40× sample into 245 μL diluent (50×). Step 3: Transfer 5 μL 2000× sample into 245 μL diluent (50×), total dilution factor = 100000×.
Guidelines for dilution: Transfer volume per step ≥ 3 μL; single-step dilution factor ≤ 100×. Mix thoroughly after each dilution and avoid bubble formation.
1. Microplate reader (capable of 450 nm absorbance measurement)
2. High-precision pipettes and matching tips: 0.5–10 μL, 5–50 μL, 20–200 μL, 200–1000 μL
3. 37°C incubator
4. Distilled or deionized water
1. Remove kit from refrigeration 10 minutes in advance and equilibrate to room temperature.
2. Serial Standard Working Solution Preparation: Reconstitute lyophilized standard with 1 mL Assay Diluent, incubate 15 minutes for full dissolution, then mix gently (stock concentration = 10 ng/mL). Prepare serial standards at concentrations: 10, 5, 2.5, 1.25, 0.625, 0.312, 0.156, 0 ng/mL. Serial dilution method: Prepare 7 EP tubes pre-filled with 500 μL Assay Diluent each. Transfer 500 μL 10 ng/mL standard into the first tube to make 5 ng/mL solution, then serially transfer 500 μL to each subsequent tube for gradient dilution. The final tube (0 ng/mL) serves as blank control with no standard added.

3. HRP-antibody Working Solution: Centrifuge 100× HRP-antibody at 1000×g for 1 minute 15 minutes before use. Dilute to 1× working concentration with Assay Diluent (example: 10 μL concentrate + 990 μL diluent). Prepare fresh right before use.
4. 1× Wash Buffer Dilution: Mix 10 mL 20× concentrated Wash Buffer with 190 mL distilled water. Crystallization of concentrated wash buffer upon refrigeration is normal; fully dissolve all crystals at room temperature before dilution.
1. Take required microplate strips from the aluminum foil pouch equilibrated at room temperature for 10 minutes. Seal unused strips in a zip bag and store at 4°C.
2. Sample Loading: Add 100 μL sample or serially diluted standard to respective wells; add 100 μL Assay Diluent to blank wells. Seal plate with sealing film and incubate at 37°C for 60 minutes. Recommendation: Dilute all test samples at minimum 1× with Assay Diluent before loading to reduce matrix interference. Multiply final calculated concentration by the dilution factor. Prepare technical replicates for all samples and standards.
3. Plate Washing: Discard liquid from wells, add 300 μL 1× Wash Buffer per well, incubate 1 minute, flick out buffer and tap plate dry on absorbent paper. Repeat wash cycle 3 times (automated plate washer is acceptable).
4. HRP-antibody Addition: Dispense 100 μL diluted HRP-antibody to each well, seal plate and incubate at 37°C for 30 minutes.
5. Plate Washing: Discard liquid, add 300 μL 1× Wash Buffer per well, incubate 1 minute, flick buffer and tap dry. Repeat wash cycle 5 times (automated plate washer is acceptable).
6. Substrate Development: Add 90 μL TMB substrate to each well, seal plate and incubate at 37°C protected from light for 15 minutes.
7. Reaction Termination & Reading: Remove microplate, add 50 μL Stop Solution to every well immediately, then measure absorbance (OD value) at 450 nm without delay.
1. Calculate the average OD value of replicate wells for standards and samples, subtract blank well OD to obtain corrected absorbance values. Plot a four-parameter logistic standard curve on a log-log graph with concentration as the X-axis and corrected OD as the Y-axis.
2. If sample OD exceeds the upper limit of the standard curve, dilute the sample appropriately and re-test; multiply the calculated concentration by the corresponding dilution factor for final reporting.
Reference Standard Curve Raw Data (For Reference Only)
|

Note: This figure is for reference only; sample concentration shall be calculated based on the standard curve generated from experimental data of each individual test run.
1. Precision: Intra-assay coefficient of variation (CV) < 10%; inter-assay CV < 10%.
2. Spike Recovery: Three concentration levels of human Lecanemab were spiked into healthy human serum, plasma and cell culture supernatant to calculate recovery rates.
|
3. Linearity upon Serial Dilution: High-concentration human Lecanemab was spiked into 4 individual batches of serum, plasma and cell culture supernatant, serially diluted within the standard curve dynamic range to assess linearity.
|
If unsatisfactory assay results occur, take photos of color development, save experimental data, retain used microplate strips and unused reagents, then contact our technical support for troubleshooting. Refer to the troubleshooting items below:
1. Poor standard curve linearity
① Incorrect standard dilution: Reconstitute and dilute standard strictly following recommended protocol.
② Inaccurate pipetting: Calibrate pipettes regularly and verify tip tightness/sealing performance.
③ Evaporation of reaction solution: Seal microplate with plate sealer film.
④ Incomplete washing: Perform sufficient wash cycles with adequate volume of wash buffer.
⑤ Foreign residues at well bottom: Wipe microplate bottom prior to absorbance measurement.
2. Faint or absent color development
① Insufficient incubation duration: Adhere to specified incubation time.
② Incorrect incubation temperature: Incubate at recommended temperature.
③ Insufficient reagent addition volume: Verify pipette accuracy and follow operating procedure strictly.
④ Improper reagent dilution: Double-check dilution steps for all reagents.
⑤ Inactivated enzyme conjugate: Mix enzyme conjugate with substrate and verify activity via color reaction test.
3. Low OD readings
① Wrong microplate reader parameter: Confirm detection wavelength setting on instrument.
② Missing stop solution addition: Add appropriate volume of stop solution.
③ Overlong delay before plate reading: Measure absorbance immediately after termination.
④ Excessively high analyte concentration in sample: Determine optimal dilution factor via preliminary pre-test.
⑤ Too low analyte concentration in sample: Determine optimal dilution factor via preliminary pre-test.
6. High background signal
① Contaminated TMB substrate: Replace with fresh substrate solution.
② Excess substrate incubation time: Strictly control color development duration.
③ Wrong dilution for detection antibody or enzyme conjugate: Dilute reagents in accordance with official recommended protocol.
④ Incomplete plate washing: Carry out adequate wash cycles with sufficient wash buffer volume.
| EJ1515725 | Components | Appearance | 48T | 96T | Storage |
| EJ1515725A | Pre-coated Assay Plate | — | 48wells | 96wells | 2-8℃. |
| EJ1515725B | Standard | Solid | 1vial | 2 vials | 2-8℃. |
| EJ1515725C | Universal Diluent | Liquid | 1×20mL | 2×20mL | 2-8℃. |
| EJ1515725D | HRP-antibody (100×) | Liquid | 60μL | 120μL | 2-8℃. |
| EJ1515725E | Wash Buffer (20×) | Liquid | 1×10mL | 2×10mL | 2-8℃. |
| EJ1515725F | TMB Substrate | Liquid | 5mL | 10mL | 2-8℃. |
| EJ1515725G | Stop Solution | Liquid | 3mL | 6mL | 2-8℃. |
| EJ1515725H | Plate Sealer | — | 4 pieces | 4 pieces | 2-8℃. |
Comprehensive hazard, handling, storage, and regulatory compliance document.
Download SDS →Lot-specific quality data. Enter your lot number to retrieve the exact COA.
Look up COA →Full quality attributes and acceptance criteria for this grade.
View spec sheet →Find and download the COA for your product by matching the lot number on the packaging.
| Lot Number | Certificate Type | Date | Item |
|---|---|---|---|
| Certificate of Analysis | Jul 14, 2026 | EJ1515725 |
Our grade selection guide covers purity, stabilizer status, and application suitability for all variants in our catalog.
View BioReagent grade guide → View for Enzyme immunoassay(ELISA) grade guide →